This page is printable and fits in 1 A4 paper.
 Cable Extension

Should it be necessary to extend the cable, EESIFLOs only recommend RG62U co-axial cable. The connection should be SOLDERED and connected as below.

The transducer cable to RG62U cable
The red cable from the transducer is soldered to the single conductor on the RG62U cable and the black and screen from the transducer are connected to the screen of the RG62U co-axial cable.

RG62U cable and input connection on the instrument
The other end of the RG62U cable should be connected to the transducer connection with the single conductor going to the red marked terminal and the screen going to the SCR marked terminal.

The extra length of cable should not exceed 250m. This distance could be shorter if the cable is run close to high voltage cables.

Copyright 2002 EESIFLO http://www.eesiflo.com