 Relay 1
The relays can be used either for an external counter, high alarm or a low alarm. The difference is that a high alarm has its reset below the setpoint, and a low alarm has its reset above the setpoint.
DEFAULT Off
 Relay 2
As above.
DEFAULT Off
 Counter 1
The external counter or sampler can be set here. Select the amount of counts before the sampler will be activated.
DEFAULT 1000
 Relay 1 High or Low depending upon selection
This is the setting for the High or the Low alarm depending on which is selected.
DEFAULT 0.00mm
 Relay 1 Reset
This is where you can enter the reset. Please note that if it is a high alarm then the reset must be below and if it is a low alarm then the reset must be above the setting.
DEFAULT 0.00mm
 Relay 2 High or Low depending upon selection
As above.
DEFAULT 0.00mm
 Relay 2 Reset
As above.
DEFAULT 0.00mm
 Totaliser Reset
The totaliser can be reset here.
DEFAULT No
 Zero Offset
Should there be a small offset at the Zero this prompt can be used to offset it.
DEFAULT No
 Offset
Up to a maximum of 50 mm minus and positive.
DEFAULT 0mm
 Use Temperature Compensation
Select Yes only if a EESIFLO10C is used or a Temperature Transducer is connected.
DEFAULT No
 Set New Password
This prompt will allow you to change the default factory password.
DEFAULT No
 Factory Reset
This prompt will reset all values entered back to factory settings including the password. Please write down all settings before using this function.
DEFAULT No