EASZ-11 was designed with a very simple configuration program.
This allows the technician to set up EASZ-11 without the
aid of a complicated source-code book. There are no references
to any codes in EASZ-11. The set up procedure is all menu
driven with the aid of questions and multiplechoice answers.
Connect up the mains and the Transducer connections as described
on the EASZ-11 board or in the EASZ-11 manual.
PLEASE NOTE! ALL CONNECTIONS ARE CAPABLE OF BEING UNPLUGGED
FROM THE PCB.
Aim the Transducer at a wall about 1.5 meters away and
check the display. It should read the following.
FL = 60.13lt/Sec
Tot 346568 |
If you press ENTER the FL (FLOW) reading will change to the
head height as below.
FL = 60.13lt/Sec
Tot 346568 |
If the reading is below 500mm then move the transducer closer
to the wall.
If the reading is above 500mm then move the transducer away
from the wall.
You may now proceed and check other distances.
PLEASE NOTE the Totaliser should be counting.
Step 1
Simply press SCROLL until the SECURITY CODE prompt
is displayed.
FL = 60.13lt/Sec
Tot 346568 |
|
|
Step 2
Enter the code 5159 by pressing the UP and DOWN keys.
Step 3
Press ENTER.
Code has been accepted.
Code not accepted, please try again.
FL = 60.13lt/Sec
Tot 346568 |
|
|
The display will indicate that it has accepted the code by
requesting the application, either mm or inch.
If you would like to carry on programming, the configuration
list is on page
5
|
|