 Setting up the Temperature Sensors

In the following steps we will "Link" the output from the two temperature sensors to the Flow Transducer installed on Channel A to produce Heat Flow Measurements.

  1. On the main menu, using the 4 and 6 keys, scroll over to the SF (Special Functions) menu and press ENTER
  2. Using the 2 and 8 keys, scroll up or down until SYSTEM Setting is displayed, then press ENTER
  3. Using the 2 and 8 keys, scroll up or down until Proc. Inputs appears then press ENTER
  4. Using the 2 and 8 keys, scroll up or down until Link temperatures appears then press ENTER
  5. The following screen will appear

    A: T-Inlet
    Input T1

The line A: T-Inlet is asking which Temperature Sensor is installed as Channel A's Inlet temperature sensor. If you've followed the above steps, this is by default Input T1

Press ENTER to accept this default value and a similar screen will appear to define Channel A's Outlet temperature sensor as follows:

A: T-Outlet
Input T2

The top line A: T-Outlet is asking which Temperature Sensor is installed as Channel A's Outlet temperature sensor. If you've followed the above steps, the default is Input T2

Press ENTER to accept this value and another screen will appear for Channel A's T3 sensor as follows:

A: T(3)
No measuring

The top line A: T(3) is asking which Temperature Sensor is installed as Channel A's (T3) temperature sensor. If you've followed the above steps, the default No Measuring should be displayed

Press ENTER to accept this value and another screen will appear for Channel A's T4 sensor as follows:

A: T(4)
No measuring

The top line A: T(4) is asking which Temperature Sensor is installed as Channel A's (T4) temperature sensor. If you've followed the above steps, the default No Measuring should be displayed

Press ENTER to accept this value and another screen will appear repeating the entire process for Channel B. All the above screens should appear for Channel B with the default Temperature Sensor of No measuring.

Press Enter to accept the No measuring inputs for all 4 of Channel B's inputs

B: T-Inlet (Outlet, T3 and T4)
No measuring

 