 Selecting Output Options >OPT<
  1. At the Main Menu use the 4 and 6 keys, scroll to OPT (Output Options) then press ENTER.
  2. Choose Channel A (or B) then press ENTER
  3. The EESIFLO 6000 will prompt you for the Physical Quantities for your desired output. Use the 2 and 8 keys to scroll up and down through the available list. Press ENTER.
  4. Using the 2 and 8 keys, choose your Velocity In units from the available list then press ENTER.
  5. Answer NO to the prompt for Temperature T1 then press ENTER
  6. Choose a Damping (averaging) time in seconds then press ENTER
  7. Answer NO to Store Measured Data then press ENTER
  8. Answer NO to Serial Output then press ENTER
  9. Enter YES to the Current Loop I1: prompt.
  10. Choose either Absolute or Signed at the Measured Values prompt to indicate if you would like only positive (Absolute) or both positive and negative (Signed) flow.
  11. At the Zero-Scale Value prompt enter a value that will represent the 4 mA output signal.
  12. At the Full-Scale Value prompt enter a value that will represent the 20 mA output signal. Once entered you will return to the Main Menu.
 Enter the Measure Mode >MEA<
  1. Using the 4 and 6 keys, scroll back to MEA (Measuring) then press ENTER
  2. Below each channel will appear a minus sign (-), a check mark () or a dot ( . ). Using the 4 and 6 keys, select Channel A (or B), then ensure there is a check mark below that channel by pressing the 2 or 8 keys. Then press ENTER
  3. A Reflection Mode question will appear. If you choose NO the EESIFLO 6000 will measure in the Direct Mode, where the transducers are mounted on opposite sides of the pipe. If you choose YES, the EESIFLO 6000 will measure in the Reflection Mode where the transducers are mounted on the same side of the pipe. It is recommended you answer YES and mount the transducers on the same side of the pipe. However, if problems are experienced getting good solid flow data, it is recommended you go through the MEA routine again, this time answering NO to this prompt, then remembering that transducers will be mounted on opposite sides of the pipe.
  4. Press ENTER and the EESIFLO 6000 will indicate how far apart the transducers must be separated. This display may indicate the values are in millimeters, when in fact they are in inches.
  5. Select an appropriate transducer Installation Site by following the guidelines in the manual pages 3-1 and 3-2
  6. Apply Ultrasonic Couplant to the bottom of the Transducers
  7. Install Transducers with arrows pointing in the same direction.
  8. Ensure Transducers are correct distance apart by pressing ENTER and reviewing the SIGNAL STRENGTH as discussed on page 5-2
  9. A GREEN signal indicates you have sufficient signal strength to begin measuring. A RED signal will indicate you should re-check your transducers installation.
  10. Press ENTER and you are now displaying and outputting the flow data you've selected !!! Once measurements are complete, press BRK 3 times to shut the instrument off.