| This
page is printable and fits in 1 A4 paper. |
|
| Selecting
Output Options >OPT< |
-
At the Main Menu use the 4 and 6
keys, scroll to OPT (Output Options) then press ENTER.
-
Choose Channel A (or B) then press
ENTER
-
The EESIFLO 6000 will prompt you
for the Physical Quantities for your desired output.
Use the 2 and 8 keys to scroll up and down through the
available list. Press ENTER.
-
Using the 2 and 8 keys, choose your
Velocity In units from the available list then press
ENTER.
-
Answer NO to the prompt for Temperature
T1 then press ENTER
-
Choose a Damping (averaging) time
in seconds then press ENTER
-
Answer NO to Store Measured Data
then press ENTER
-
Answer NO to Serial Output then press
ENTER
-
Enter YES to the Current Loop I1:
prompt.
-
Choose either Absolute or Signed
at the Measured Values prompt to indicate if you would
like only positive (Absolute) or both positive and negative
(Signed) flow.
-
At the Zero-Scale Value prompt enter
a value that will represent the 4 mA output signal.
-
At the Full-Scale Value prompt enter
a value that will represent the 20 mA output signal.
Once entered you will return to the Main Menu.
| Enter
the Measure Mode >MEA< |
-
Using the 4 and 6 keys, scroll back
to MEA (Measuring) then press ENTER
-
Below each channel will appear a
minus sign (-), a check mark (  )
or a dot ( . ). Using the 4 and 6 keys, select Channel
A (or B), then ensure there is a check mark below that
channel by pressing the 2 or 8 keys. Then press ENTER
-
A Reflection Mode question will
appear. If you choose NO the EESIFLO 6000 will measure
in the Direct Mode, where the transducers are mounted
on opposite sides of the pipe. If you choose YES, the
EESIFLO 6000 will measure in the Reflection Mode where
the transducers are mounted on the same side of the
pipe. It is recommended you answer YES and mount the
transducers on the same side of the pipe. However, if
problems are experienced getting good solid flow data,
it is recommended you go through the MEA routine again,
this time answering NO to this prompt, then remembering
that transducers will be mounted on opposite sides of
the pipe.
-
Press ENTER and the EESIFLO 6000
will indicate how far apart the transducers must be
separated. This display may indicate the values are
in millimeters, when in fact they are in inches.
-
Select an appropriate transducer
Installation Site by following the guidelines in the
manual pages 3-1 and 3-2
-
Apply Ultrasonic Couplant to the
bottom of the Transducers
-
Install Transducers with arrows
pointing in the same direction.
-
Ensure Transducers are correct distance
apart by pressing ENTER and reviewing the SIGNAL STRENGTH
as discussed on page 5-2
-
A GREEN signal indicates you have
sufficient signal strength to begin measuring. A RED
signal will indicate you should re-check your transducers
installation.
-
Press ENTER and you are now displaying
and outputting the flow data you've selected !!! Once
measurements are complete, press BRK 3 times to shut
the instrument off.
|
|