 Troubleshooting guide

PROBLEM POSSIBLE SOLUTION
Meter reading lower than expected
  • Source particles velocity not indicative of average velocity
  • Relocate sensor to a position where source
        particles are expected to be moving at the average
        velocity
  • Incorrect mounting of flow sensor
  • Remount sensor correctly
  • Programming error
  • Review all programmed entries
  • Flow rate lower than expected
  • Investigate possible causes and confirm flow rate
        independently
  • Insufficient particle size or concentration
  • Locate sensor at position where acceptable particle
        size or concentration is expected.
  • Inject air into the line
  • Meter reading when there is no flow
  • Local ultrasonic noise source
  • Relocate sensor or remove noise source
  • “Poor signal” displayed when flow exists
  • Insufficient particle size or concentration
  • Locate sensor at position where acceptable particle
        size or concentration is expected.
  • Inject air into the line
  • Sensor coupling to pipe poor
  • Remount sensor to pipe correctly
  • Meter reading higher than expected
  • Programming error
  • Review all programmed entries
  • Flow rate higher than expected
  • Investigate possible causes and confirm flow rate
        independently
  • Particle velocity at sensor not indicative of average velocity
  • Relocate sensor to a position where source
        particles are expected to be moving at the average
        velocity
  • Incorrect mounting of flow sensor
  • Remount sensor correctly
  • Local electrical noise
  • Relocate sensor
  • Meter reading erratic
  • Particle velocity at sensor not indicative of average velocity and
        erratic
  • Relocate sensor to a position where the velocity
        profile is expected to be suitable