 Battery

Recharging and battery care

  • The10P is supplied with 4 x 1.2V 650mAh AA size Ni-Cd removable cells and a 110/230Vac battery charger.
  • Charge battery fully before first use and thereafter recharge only when fully discharged. The 10P will indicate an error message when battery charge is low and automatically switch off.
  • The 110/230Vac battery charger supplied as standard with the 10P is a 65mA constant current charger. The 10P should be switched off during charging. The charge time for the 650mAh cells supplied is 14-16 hours (maximum 20 hours). THE CELLS MUST NOT BE OVER CHARGED. Under charging of the cells will reduce the life and capacity of the cells.

Low power consumption

  • The10P is designed for low power consumption allowing over 10 hours operation before re-charging the battery. The 10P features a Low Battery warning with automatic power down.

Storage

  • If the 10P is to be stored for an extended time period the cells should be removed.
 Menu System (10P Version 1.00)

The 10P menu system is easy to use and designed for programming simplicity.

With the 10P powered up the 10P will test the suitability of the flow signal. If the signal is suitable the flow total and flow rate are displayed, if not an error message is displayed.

  00000000 lt
3.9768 l/s
 
START PROGRAMMING - Press "MENU"

Units
  Mn_1 units?
Metric
 
Metric and English units of rate and total measurement are available.
Press until desired value is displayed and MENU to continue.

Pipe ID
  Mn_2 Pipe ID mm
53.4
 
The precise dimension of the pipe internal diameter (ID) at the point of measurement must be entered.
Use the button to locate the curser below the number to be changed and press the button until the desired value is displayed and MENU to continue

Rate units
  Mn_3 rate units?
l/s
 
Press until desired unit is displayed and MENU to continue.